Laboratory Equipment > Analytical Balances > Semi Micro - 0.01 mg | ||||||
|
||||||
OverviewIntuitive to operate with robust construction. 92 g capacity; 0.01 mg readability; large, bright display; internal adjustment; easy leveling; chemical resistance; USB and RS232 connection; passcode protection
Capacity x Readability
Manufature:
Mettler Toledo92 g x 0.01 mg Pan Size 80 mm (Diameter) SKU: 30706701 Weight: 22.00 lb |
MA Analytical Balances
You can rely on accurate results with these high-quality balances that are intuitive to operate and easy to clean. Built-in applications provide simple instructions, while automatic calculations and electronic data transfer minimize errors and make documentation easy.
The bright LCD screen provides excellent readability, while the glass surface resists wear and tear and is easy to clean. The touch keys make it simple to operate the balance and navigate through the menus in the user interface.
Accessories
Enhance performance, improve ergonomics, and handle your data efficiently with our wide range of accessories, including EasyDirect Balance data management software, printers, weights, and density kit
Features
Performance
Efficient Operation
Quality Assurance
Data Management
Sustainable Value
Model | MA55 30666316 |
MA95 30706701 |
MA155DU 30697382 |
MA54 30697385 |
MA104 30697408 |
MA204 30697411 |
|||
---|---|---|---|---|---|---|---|---|---|
Capacity x Readability | 52 g x 0.01 mg | 92 g x 0.01 mg | 62 g x 0.01 mg 152 g x 0.1 mg |
52 g x 0.1 mg | 120 g x 0.1 mg | 220 g x 0.1 mg | |||
Legal for Trade Model | MA55/A 30666317 |
MA95/A 30706702 |
MA155DU/A 30697383 |
MA54/A 30697386 |
MA104/A 30697409 |
MA204/A 30697412 |
|||
CC#24-011 Class I | |||||||||
52 g x 1 mg | 92 g x 1 mg | 152 g x 1 mg | 52 g x 1 mg | 120 g x 1 mg | 220 g x 1 mg | ||||
Weighing Pan Diameter | 80 mm | 80 mm | 80 mm | 90 mm | 90 mm | 90 mm | |||
Repeatability (at 5% load) | 0.03 mg | 0.03 mg | 0.03 mg | 0.1 mg | 0.1 mg | 0.1 mg | |||
Linearity deviation | 0.1 mg | 0.1 mg | 0.2 mg | 0.2 mg | 0.2 mg | 0.2 mg | |||
Sensitivity offset (at nominal load) | 0.4 mg | 0.4 mg | 0.8 mg | 0.3 mg | 0.5 mg | 0.8 mg | |||
Typical values | |||||||||
Repeatability (at 5% load) | 0.015 mg | 0.015 mg | 0.015 mg | 0.08 mg | 0.08 mg | 0.08 mg | |||
Linearity deviation | 0.03 mg | 0.06 mg | 0.06 mg | 0.06 mg | 0.06 mg | 0.06 mg | |||
Sensitivity offset (at nominal load) | 0.25 mg | 0.25 mg | 0.5 mg | 0.15 mg | 0.3 mg | 0.5 mg | |||
Minimum weight (USP, tolerance = 0.10%) | 30 mg | 30 mg | 30 mg | 160 mg | 160 mg | 160 mg | |||
Minimum weight (tolerance = 1%) | 3 mg | 3 mg | 3 mg | 16 mg | 16 mg | 16 mg | |||
Settling time | 4 s | 4 s | 4 s | 16 s | 16 s | 16 s | |||
Dimensions and other specifications | |||||||||
Data Management | USB-A, RS232, Bluetooth (optional) | ||||||||
Adjustment | Internal adjustment | ||||||||
Display | LCD hybrid touchscreen | ||||||||
Balance dimensions (W × D × H) | 209 × 354 × 354 mm |
|